For one-time use by medical units during uterine and salpingography examinations of patients.
The product is composed of a guide head, a balloon, a tube body, a filling chamber, a valve, an injection chamber, an outlet hole and a balloon filling device. The balloon refiller is a standard screw syringe containing a quantity of sterile water or 0.9% saline for balloon refilling.
Product Number |
Size |
Balloon capacity |
Length (mm) |
WD-010-07 |
7Ch/Fr |
1.5ml |
360 |
Features :
1, silicone uterine angiography catheter is specially designed and manufactured for latex, PVC and other biocompatibility is poor, the product is easy to age, strong irritation shortcomings, can effectively improve the use of comfort and safety. Safe, non-toxic, transparent and smooth.
2. The balloon wall is thin, strong and smooth, and the connection between the balloon and the tube body is smooth and transitive, which can reduce the risk of cavity trauma during intubation or extubation, and improve the comfort of patients.
3, 100% pure silicone manufacturing, transparent, soft, smooth, comfortable, non-toxic medical silicone material.
4, is through the cervical tube into the uterine cavity composed of high specific gravity of the substance iodine, X-ray and surrounding tissue formed obvious artificial contrast, so that the lumen appears, so as to understand the uterus and fallopian tube lumen conditions. Angiography can not only indicate whether the fallopian tube is unobstructed, the location of obstruction, but also observe the shape of the uterine cavity. This kind of examination is relatively safe, does not require anesthesia, no obvious pain, and can be accepted by patients.
5, 1.5ml capacity balloon, "X "development line, can observe the status of the catheter in the body under the imaging equipment.
6, with nickel-titanium alloy guide wire, strengthen the hardness of the catheter to facilitate the introduction of fallopian tube cavity.
Packaging requirements :
1, the pipe body transparent without impurities, no stains, no cracking, accessories connected in place without cracking, the balloon transition smooth without segment difference, the accessories are firmly connected and smooth.
2, the model specification is clearly marked, the overall smooth without obvious burrs.
3. Its inner packaging is first packed with mesh, then transparent domestic glue-coated paper +PET/PE4.7C paper plastic bag, the middle packaging is made of white cardboard box, and the outer packaging is made of K=A reinforced five-layer corrugated box.
4. The product is sterilized by ethylene oxide and provided in sterile form.
Instructions for use:
1. Tear the single package from the "opening" to take out the catheter, insert the lubricated catheter through the patient's vagina into the uterine cavity, and ensure that the balloon is completely entered into the uterine cavity.
2. Open the double-pass valve, inject sterile water or normal saline into the balloon, and close the double-pass valve after the balloon is filled.
3. Uterus and salpingography were performed after contrast agent was injected into the injection cavity.
4. After the inspection is completed, the catheter can be pulled out after draining the liquid in the balloon.
Precautions:
1, patients with allergic history of silicone products should use with caution.
2. The product is valid for three years from the date of sterilization. Recognize the "expiration date" before use, and do not use after expiration.
3. The screw syringe equipped with the product contains a quantity of sterile water or 0.9% concentration of normal saline, which is used for balloon filling without additional extraction.
4. this product is only for one-time use, destroy after use.
5. avoid light, dry, heat storage and transportation.
6. this product is only for the use of professionally trained medical personnel.
7. the package is damaged, do not use. 8. The product can be used after seven days of sterilization.